Abstract / Description of output
Although CCD X-ray detectors can be faster to use, their large-area versions can be much more expensive than similarly sized photographic plate detectors. When indexing X-ray diffraction patterns, large-area detectors can prove very advantageous as they provide more spots, which makes fitting an orientation easier. On the other hand, when looking for single crystals in a polycrystalline sample, the speed of CCD detectors is more useful. A new setup is described here which overcomes some of the limitations of limited-range CCD detectors to make them more useful for indexing, whilst at the same time making it much quicker to find single crystals within a larger polycrystalline structure. This was done by combining a CCD detector with a six-axis goniometer, allowing the compilation of images from different angles into a wide-angled image. Automated scans along the sample were coupled with image processing techniques to produce grain maps, which can then be used to provide a strategy to extract single crystals from a polycrystal.
Original language | English |
---|---|
Pages (from-to) | 1342-1345 |
Number of pages | 4 |
Journal | Journal of Applied Crystallography |
Volume | 48 |
DOIs | |
Publication status | Published - Aug 2015 |
Keywords / Materials (for Non-textual outputs)
- laboratory-based Laue X-ray diffraction
- grain mapping