Calibration of Photomultiplier Arrays

F. Neves, V. Chepel, D. Yu. Akimov, H. araujo, E. barnes, V. belov, A. burenkov, B. Currie, B. Edwards, C. Ghag, M. Horn, A. hughes, G. kalmus, A. kobyakin, A. kovalenko, V. lebedenko, A. Lindote, M. lopes, R. Luscher, K. LyonsP. Majewski, Alex Murphy, J. Pinto da Cunha, R. Preece, J. quenby, P. scovell, C. Silva, V. solovov, N. smith, P. smith, V. stekhanov, T. sumner, C. Thorne, R. walker

Research output: Contribution to journalArticlepeer-review

Abstract / Description of output

A method is described that allows calibration and assessment of the linearity of response of an array of photomultiplier tubes. The method does not require knowledge of the photomultiplier single photoelectron response model and uses science data directly, thus eliminating the need for dedicated data sets. In this manner all photomultiplier working conditions (e.g. temperature, external fields, etc) are exactly matched between calibration and science acquisitions. This is of particular importance in low background experiments such as ZEPLIN-III, where methods involving the use of external light sources for calibration are severely constrained.
Original languageUndefined/Unknown
Pages (from-to)13-18
JournalAstroparticle Physics
Volume33
Issue number1
DOIs
Publication statusPublished - Feb 2010