TY - JOUR
T1 - On-site monitoring of trace levels of free manganese in sea water via sonoelectroanalysis using a boron-doped diamond electrode
AU - Goodwin, A.
AU - Lawrence, A.L.
AU - Banks, C.E.
AU - Wantz, F.
AU - Omanović, D.
AU - Komorsky-Lovrić, Š.
AU - Compton, R.G.
AU - Lawrence, Andrew
PY - 2005/3/28
Y1 - 2005/3/28
N2 - Cathodic stripping voltammetry at a boron-doped diamond electrode utilising an insonated accumulation protocol is shown to be possible for the on-site trace detection of free manganese in seawater samples. No sample pre-treatment is required with the application of ultrasound providing a sensitive and selective protocol for the analysis of manganese. The protocol is used in the evaluation of manganese in seawater samples taken near the site of a former ferromanganese factory in the town Šibenik, Croatia.
AB - Cathodic stripping voltammetry at a boron-doped diamond electrode utilising an insonated accumulation protocol is shown to be possible for the on-site trace detection of free manganese in seawater samples. No sample pre-treatment is required with the application of ultrasound providing a sensitive and selective protocol for the analysis of manganese. The protocol is used in the evaluation of manganese in seawater samples taken near the site of a former ferromanganese factory in the town Šibenik, Croatia.
UR - http://www.scopus.com/inward/record.url?eid=2-s2.0-25444483843&partnerID=8YFLogxK
U2 - 10.1016/j.aca.2004.11.009
DO - 10.1016/j.aca.2004.11.009
M3 - Article
AN - SCOPUS:25444483843
VL - 533
SP - 141
EP - 145
JO - Analytica Chimica Acta
JF - Analytica Chimica Acta
SN - 0003-2670
IS - 2
ER -