Abstract / Description of output
A sensitive liquid chromatography-mass spectrometric method was validated for the quantification of RU38486 (mifepristone) in human and murine plasma. The analyte and internal standard (alfaxolone) were extracted by liquid-liquid extraction with diethyl ether, resolved on a C18 column using gradient elution with methanol and ammonium acetate and detected after positive electrospray ionization (m/z 430-->372; m/z 333-->297, respectively). Quantification was linear over the range 0.5-500ng (r(2)>0.997), precise and accurate (intra-assay RSD< or =7.2%, RME< or =8.2%; inter-assay RSD< or =15.7% RME< or =10.2%). The limit of quantification (LOQ) was 50pg injected on column, permitting reproducible analysis of RU38486 in small volumes of plasma.
Original language | English |
---|---|
Pages (from-to) | 497-501 |
Number of pages | 5 |
Journal | Journal of Chromatography B |
Volume | 877 |
Issue number | 5-6 |
DOIs | |
Publication status | Published - Feb 2009 |
Keywords / Materials (for Non-textual outputs)
- RU38486
- Mifepristone
- Steroid
- Mass spectrometry
- Liquid chromatography