Time multiplexed active neural probe with 678 parallel recording sites

Bogdan C Raducanu, Refet F Yazicioglu, Carolina M Lopez, Marco Ballini, Jan Putzeys, Shiwei Wang, Alexandru Andrei, Marleen Welkenhuysen, Nick Van Helleputte, Silke Musa, Robert Puers, Fabian Kloosterman, Chris Van Hoof, Srinjoy Mitra

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

We present a high density CMOS neural probe with active electrodes (pixels), consisting of dedicated in-situ circuits for signal source amplification. The complete probe contains 1356 neuron sized (20×20 μm 2 ) pixels densely packed on a 50 μm thick, 100 μm wide and 8 mm long shank. It allows simultaneous high-performance recording from 678 electrodes and a possibility to simultaneously observe all of the 1356 electrodes with increased noise. This considerably surpasses the state of the art active neural probes in electrode count and flexibility. The measured action potential band noise is 12.4 μVrms, with just 3 μW power dissipation per electrode amplifier and 45 μW per channel (including data transmission).
Original languageUndefined/Unknown
Title of host publication2016 46th European Solid-State Device Research Conference (ESSDERC)
PublisherInstitute of Electrical and Electronics Engineers
Pages385-388
Number of pages4
ISBN (Print)9781509029693
DOIs
Publication statusPublished - 20 Oct 2016
Event2016 46th European Solid-State Device Research Conference (ESSDERC) -
Duration: 12 Sept 201615 Sept 2016

Conference

Conference2016 46th European Solid-State Device Research Conference (ESSDERC)
Period12/09/1615/09/16

Keywords / Materials (for Non-textual outputs)

  • active