Use of a confocal optical device for centring a diamond anvil cell in single-crystal X-ray diffraction experiments

Chengdao Hu, Cameron j. g. Wilson, Daniel m. Scully, Tobias Stuerzer, Simon Parsons

Research output: Contribution to journalArticlepeer-review

Abstract

High-pressure crystallographic data can be measured using a diamond anvil cell (DAC), which allows the sample to be viewed only along a cell vector which runs perpendicular to the diamond anvils. Although centring a sample perpendicular to this direction is straightforward, methods for centring along this direction often rely on sample focusing, measurements of the direct beam or short data collections followed by refinement of the crystal offsets. These methods may be inaccurate, difficult to apply or slow. Described here is a method based on precise measurement of the offset in this direction using a confocal optical device, whereby the cell centre is located at the mid-point of two measurements of the distance between a light source and the external faces of the diamond anvils viewed along the forward and reverse directions of the cell vector. It is
shown that the method enables a DAC to be centred to within a few micrometres
reproducibly and quickly.
Original languageEnglish
JournalJournal of Applied Crystallography
Volume57
Issue number5
Early online date20 Sept 2024
DOIs
Publication statusE-pub ahead of print - 20 Sept 2024

Fingerprint

Dive into the research topics of 'Use of a confocal optical device for centring a diamond anvil cell in single-crystal X-ray diffraction experiments'. Together they form a unique fingerprint.

Cite this