Validation of a new non-invasive automatic monitor of respiratory rate for postoperative subjects

G. B. Drummond, A. Bates, J. Mann, D. K. Arvind

Research output: Contribution to journalArticlepeer-review

Abstract

Background Respiratory rate is an important measurement in patient care but frequently poorly assessed. We set out to develop a simple non-invasive device to reliably measure respiratory movements and estimate respiratory rate, in clinical circumstances.Methods Respiratory movement was detected with an encapsulated tri-axial accelerometer (Orient speck) and the data transmitted wirelessly to a computer for analysis. We studied subjects after gynaecological surgery who received opioid analgesia, and compared the derived signal with a signal from nasal cannula using directly matched breaths and within the same 5 min epoch. We analysed the signals for 5 min epochs over a 15 h recording period.Results For matched breath analysis, the instantaneous respiratory rates matched within 2 bpm on 86% of occasions. A similar match was found between epoch averages of the respiratory rate. The mean absolute difference between the respiratory rate measured by nasal cannula and Orient speck was 0.6 bpm. The Orient speck generated reliable measures of respiratory rate every 5 min in 95.4% of epochs.Conclusions The Orient speck provides a reliable measure of respiratory rate at frequent intervals in subjects receiving patient-controlled morphine analgesia after surgery.
Original languageEnglish
Pages (from-to)462-469
Number of pages8
JournalBritish Journal of Anaesthesia
Volume107
Issue number3
DOIs
Publication statusPublished - Sept 2011

Keywords / Materials (for Non-textual outputs)

  • monitoring, physiologic
  • signal processing, computer-assisted
  • respiratory rate

Fingerprint

Dive into the research topics of 'Validation of a new non-invasive automatic monitor of respiratory rate for postoperative subjects'. Together they form a unique fingerprint.

Cite this