Edinburgh Research Explorer

Real-time measurement of tumour hypoxia using an implantable microfabricated oxygen sensor

Research output: Contribution to journalArticle

Original languageEnglish
JournalSensing and Bio-Sensing Research
Early online date11 Aug 2020
DOIs
Publication statusE-pub ahead of print - 11 Aug 2020

Abstract

Hypoxia commonly occurs within tumours and is a major cause of radiotherapy resistance. Clinical outcomes could be improved by locating and selectively increasing the dose delivered to hypoxic regions. Here we describe a miniature implantable sensor for real-time monitoring of tissue oxygenation that could enable this novel treatment approach to be implemented. The sensor uses a solid-state electrochemical cell that was microfabricated at wafer level on a silicon substrate, and includes an integrated reference electrode and electrolyte membrane. It gave a linear response to oxygen concentration, and was unaffected by sterilisation and irradiation, but showed susceptibility to biofouling. Oxygen selectivity was also evaluated against various clinically relevant electroactive compounds. We investigated its robustness and functionality under realistic clinical conditions using a sheep model of lung cancer. The sensor remained functional following CT-guided tumour implantation, and was sufficiently sensitive to track acute changes in oxygenation within tumour tissue.

Download statistics

No data available

ID: 159226832